Precautions for use and maintenance of 9" Industrial thermometer
Aug 30, 2020|
View:14271.<a href="https://www.czkb.net/Product/9"-Industrial-thermometer.html" target="_self" _href="https://www.czkb.net/Product/9"-Industrial-thermometer.html" textvalue="9" Industrial thermometer">9" Industrial thermometer should be used in an environment where the ambient air temperature is -10oC- +55oC and the relative humidity is not more than 95%.
2. The thermometer should be installed vertically on the mounting plate without vibration. (No more than 30Hz)
3. The temperature sensor must be completely immersed in the measured medium, and the temperature sensor should be inserted as deep as possible.
4. When installing, the capillary tube should be straightened and fixed with a rolling head at a distance of no more than 300 mm.
5.9" The regular working temperature of the industrial thermometer should be 1/2-3/4 of the measuring range.
6. Under any circumstances (whether it is transported, installed or used), vibration, collision and impact should be avoided for the <a href="https://www.czkb.net/product.html" target="_self" _href="https://www.czkb.net/product.html" textvalue="9" Industrial thermometer">9" Industrial thermometer.