Your location:Home   News   Company News
Two measurement errors of capillary thermometer
 Oct 19, 2020|View:1090

1. High temperature range

The temperature measurement range of capillary thermometer is 200~650℃ (also clearly recorded in the encyclopedia), and the allowable error is about 1% of the scale range. In fact, this is not right, because this kind of metal sheet is made into a spiral-shaped heat conduction sensor, which is easy to fatigue at a high temperature of 650℃, and the temperature difference is very large. Therefore, when the temperature exceeds 550℃, a thermocouple should be considered to solve this problem.

Two measurement errors of Capillary Thermometer

2. Low temperature range

It is clearly stated in the relevant information that the temperature range of capillary thermometer is 200~650℃ (this is not correct), and the low temperature can be measured to -40℃. In fact, this is not much, because the temperature calibration of capillary thermometer in the production is usually -20℃ (if low temperature is needed, if not, we usually mark 0℃ to the upper limit of the temperature required by the customer. Users in Beijing Fang District Sometimes -40°C is used, which can be done, but in the case of -40°C, it is necessary to consider whether the error can be accepted.

The difference between oil-filled and non-oil-filled capillary thermometer is mainly the difference between the two applications. Capillary thermometer is mainly used in places where the environment is vibrated.